Introduction

++

Objectives

  1. Discuss the gas delivery patterns during volume-controlled ventilation (VCV) and pressure-controlled ventilation (PCV).

  2. Describe the effect of varying rise time and inspiratory cycle criteria during pressure support ventilation and PCV.

  3. Describe how an end-inspiratory plateau can be achieved with PCV.

  4. Describe approaches to monitor gas delivery during PCV and VCV.

  5. Contrast the advantages and disadvantages of PCV and VCV.

++

Controversy has always followed the introduction of new modes of ventilation. In the late 1970s it was assist/control (continuous mandatory ventilation [CMV]) versus intermittent mandatory ventilation (IMV). In the mid- to late-1980s, it was IMV versus pressure support ventilation. A debate today is whether gas delivery should be volume-controlled or pressure-controlled.

++

Volume-Controlled Ventilation

++

All of these first-generation ICU ventilators only provided VCV, and until the 1970s, it was without the option for patient-triggered breaths. Pressure-limited ventilators (eg, Bird and Puritan-Bennett machines) have been available since the 1950s, but they were not designed for continuous ventilatory support.

++

With VCV, the variable that is constant during each breath is tidal volume. With this approach, there is a variable inspiratory pressure. With changes in respiratory mechanics (eg, resistance, compliance) and patient's effort, airway pressure varies because volume delivery is constant. With VCV, the clinician sets tidal volume (VT), flow pattern, peak inspiratory flow, rate, and trigger sensitivity. In some ventilators, inspiratory time, minute volume, and I:E ratio are set instead of VT and flow. In other ventilators, both inspiratory time and flow are set; if the inspiratory time is longer than the time required to deliver the VT, an end-inspiratory pause will result. In practice, modern ventilators provide VCV by controlling the inspiratory flow. VCV can be applied as CMV or as synchronized IMV (CMV-IMV).

++

Pressure-Controlled Ventilation

++

With pressure-controlled ventilation (PCV), a fixed pressure is applied during the inspiratory phase. In addition, inspiratory time, or I:E ratio, and trigger sensitivity are set. As respiratory mechanics (eg, resistance, compliance) and patient's effort change, VT must vary because pressure is constant. As noted in Table 7-1, the primary difference between these VCV and PCV is a fixed VT or a fixed peak inspiratory pressure (PIP), respectively. In newer generation ventilators, the clinician may also set the rise time, which is the time required for the set pressure to be reached. This occurs by varying the slope of the flow increase from baseline to peak flow.

++
Table Graphic Jump Location
Table 7-1PCV Versus VCV
++

Pressure Support Ventilation

++

Pressure support ventilation (PSV) is ...

Want remote access to your institution's subscription?

Sign in to your MyAccess profile while you are actively authenticated on this site via your institution (you will be able to verify this by looking at the top right corner of the screen - if you see your institution's name, you are authenticated). Once logged in to your MyAccess profile, you will be able to access your institution's subscription for 90 days from any location. You must be logged in while authenticated at least once every 90 days to maintain this remote access.

Ok

About MyAccess

If your institution subscribes to this resource, and you don't have a MyAccess profile, please contact your library's reference desk for information on how to gain access to this resource from off-campus.

Subscription Options

AccessAnesthesiology Full Site: One-Year Subscription

Connect to the full suite of AccessAnesthesiology content and resources including procedural videos, interactive self-assessment, real-life cases, 20+ textbooks, and more

$995 USD
Buy Now

Pay Per View: Timed Access to all of AccessAnesthesiology

24 Hour Subscription $34.95

Buy Now

48 Hour Subscription $54.95

Buy Now

Pop-up div Successfully Displayed

This div only appears when the trigger link is hovered over. Otherwise it is hidden from view.